in vivo assessment of stent recoil of biodegradable polymer-coated cobalt–chromium sirolimus-eluting coronary stent system

Clicks: 166
ID: 202533
2012
Introduction: Immediate and acute stent recoil has been observed following balloon deflation in normal and diseased coronary arteries, and the degree varies by stent design. Methods: A total of 19 patients, who underwent elective stent implantation for single de novo native coronary artery lesions, were enrolled: all patients treated with the biodegradable polymer-coated sirolimus-eluting cobalt–chromium coronary stent system (Supralimus-Core®). The immediate, acute and cumulative stent recoil was assessed by quantitative coronary angiography. The cumulative stent recoil was measured at 24 h of stent implantation. Results: The absolute late loss due to recoil was found 0.08 ± 0.19 mm for Immediate Stent Recoil (ISR), 0.05 ± 0.21 mm for Acute Stent Recoil (ASR) and 0.11 ± 0.25 mm for Cumulative Stent Recoil (CSR) respectively. Conclusions: In vivo acute stent recoil of the Supralimus-Core® has higher radial strength compared to other available standard drug-eluting stents.
Reference Key
abhyankar2012indianin vivo Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors ;Atul D. Abhyankar;Ashok S. Thakkar
Journal renewable energy
Year 2012
DOI 10.1016/j.ihj.2012.07.005
URL
Keywords

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.